http://hdl.handle.net/1893/12178
Appears in Collections: | Faculty of Health Sciences and Sport Journal Articles |
Peer Review Status: | Refereed |
Title: | Elimination of electrically induced iontophoretic artefacts: Implications for non-invasive assessment of peripheral microvascular function |
Author(s): | Ferrell, William R Ramsay, Jane Elizabeth Brooks, Naomi Lockhart, John C Dickson, Sylvia McNeece, Grainne M Greer, Ian A Sattar, Naveed |
Contact Email: | n.e.brooks@stir.ac.uk |
Keywords: | Iontophoresis Laser Doppler imaging Galvanic response Electrically induced hyperaemia Endothelium Acetylcholine Sodium nitroprusside Prostaglandins Vasodilatation Resistance Muscle strength |
Issue Date: | Sep-2002 |
Date Deposited: | 22-Apr-2013 |
Citation: | Ferrell WR, Ramsay JE, Brooks N, Lockhart JC, Dickson S, McNeece GM, Greer IA & Sattar N (2002) Elimination of electrically induced iontophoretic artefacts: Implications for non-invasive assessment of peripheral microvascular function. Journal of Vascular Research, 39 (5), pp. 447-455. https://doi.org/10.1159/000064515 |
Abstract: | Iontophoretic assessment of skin microvascular function is complicated by the occurrence of electrically induced hyperaemia, especially at the cathode. Studies were performed to identify means of reducing such effects. Skin vasodilator responses were measured using a laser Doppler imager that controlled iontophoretic current delivery. A novel feature involved monitoring voltage across the iontophoresis chambers. Comparison between responses to vehicle (distilled H2O), acetylcholine (ACh) and sodium nitroprusside (SNP) showed electrically induced hyperaemia at the cathode associated with the vehicle, whose time course overlapped with that of the SNP response. Voltage across the chambers containing drugs dissolved in H2O was significantly (p = 0.018, n = 7) lower than the voltage profile of H2O alone. H2O iontophoresis was associated with cathodal hyperaemic responses in most subjects, whereas a 0.5% NaCl vehicle produced lower voltages and eliminated this artefact. Voltage·time integral rather than charge was the prime determinant of electrically induced hyperaemic responses. No significant correlation was found between skin fold thickness and either calculated skin resistance (r2 = 0.0002) or vascular response to ACh (r2 = 0.13). Smaller chamber size led to higher voltages and greater electrically induced hyperaemic responses. These appear to be prostaglandin dependent as they were ablated by cyclooxygenase inhibition. Use of a low-resistance vehicle combined with larger chamber sizes and lower currents can prevent such artefacts, thereby increasing the robustness of this methodology for clinical assessment of endothelial function. |
DOI Link: | 10.1159/000064515 |
Rights: | The publisher does not allow this work to be made publicly available in this Repository. Please use the Request a Copy feature at the foot of the Repository record to request a copy directly from the author. You can only request a copy if you wish to use this work for your own research or private study. |
Licence URL(s): | http://www.rioxx.net/licenses/under-embargo-all-rights-reserved |
File | Description | Size | Format | |
---|---|---|---|---|
Brooks_2002_Elimination_of_Electrically_Induced_Iontophoretic_Artefacts.pdf | Fulltext - Published Version | 308.99 kB | Adobe PDF | Under Permanent Embargo Request a copy |
Note: If any of the files in this item are currently embargoed, you can request a copy directly from the author by clicking the padlock icon above. However, this facility is dependent on the depositor still being contactable at their original email address.
This item is protected by original copyright |
Items in the Repository are protected by copyright, with all rights reserved, unless otherwise indicated.
The metadata of the records in the Repository are available under the CC0 public domain dedication: No Rights Reserved https://creativecommons.org/publicdomain/zero/1.0/